SD DUAL   DUAL CHANNEL MINIATURE SPECTROMETER

The SD DUAL spectrograph has two independent spectrographs sharing the same housing. The focal planes of the Spectrographs are matched to provide simultaneous spectrum detection at a 27mm-diode array. The SD DUAL has both high quality optics and proven electronics . The optical system uses a special grating to cheive a dual channel device. Both channel outputs are detected by a highly sensitive CCD detector or low noise photodiode array. Both channels cover the same spectral range and are used in situations where a reference channel is required. 

 

Available in four spectral ranges from UV to NIR.

USB, RS232 or parallel printer port PC interface.

Complete, plug into PC and run. No plugin PC card required.

MODEL

DB-UV

DB-CL clinical

DB-CO colour

DB-NIR

Spectral Range (nm)

190 .. 350

340 .. 630

390 .. 720

600 .. 1100

Pixel Resolution (nm)

3.0

8-10

8-10

15.0

Straylight

< 0.1%

< 0,1 %

< 0,1 %

< 0,1 %

Detector

Available with low cost CCD (not UV) or Photodiode array

Optical input

2 X Optical slit accessed by SMA connectors

Digital Resolution

12 or 16 bit

Data Storage

512 kbyte (512scans)         optionally 1 Mbyte (1024 scans)

Scan rate

64 Scans/Sec 16 bit                 1000 scans/sec 12 bit

Interface

Parallel port 2Mbytes/sec          R232 9,600…115,200 baud USB 12 Mbit/s

Size

W63 X L95 X H65 mm

Detectors used:

CCD array with 2048 pixels for low cost and very good sensitivity.

Photodiode array with 256 or 512 low noise 16 bit versions. 

[Description of the Cronin Camera Electronics]